Zirconia oxygen sensor for SO-E2-960 oxygen generator
Brand: Austrian Sensore
Product Category: Zirconia Sensor
SO-E2-960 working principle of zirconia oxygen sensor for 1. oxygen generator:
BecauseThe carriers for the current in the zirconia electrolyte are oxygen ions, so when a voltage is applied to the zirconia cell, oxygen is drawn through the zirconia disk to the anode. If a perforated cover is added to the cathode of the electrolytic cell, the oxygen flows to the cathode. The rate will be limited. Limited by this rate, the current in the cell will reach saturation as the applied voltage is gradually increased. This saturation current is called the limit current, which is concentrated with oxygen in the surrounding environment. proportional to the degree.
SO-E2-960 of Zirconia Oxygen Sensor for 2. Oxygen Making MachineApplication:
Medical: oxygen concentrator, Constant temperature box, oxygen generator
Laboratory: inert gas treatment cabinet (glove box), bacteria incubator
Food industry: packaging, food inspection, Monitor fruit ripening process(Storage/Transportation)
Family/Cooking: Automated baking/Baking (high temperature>100℃)
Measurement technology: fixed/Portable oxygen meter, Measurement, air conditioning and circulation with controlled oxygen content
security technology/Monitoring: fire protection (nitrogen addition, e.g. server room), greenhouses, wine cellars, gas storage, refineries, diving, fermentation units
Electrical industry: inert gas processors and cabinets, Inert gas welding monitoring, Storage with increased nitrogen (oxidation protection), drying equipment, nitrogen concentrator, exhaust gas measurement
SO-E2-960 of Zirconia Oxygen Sensor for 3. Oxygen Making MachineAdvantages of oxygen sensors:
Wide measurement range,1%~96%Oxygen
High precision
Multiple models have linear characteristics
Small temperature dependence of sensor signal
Low cross sensitivity
Long service life
In most cases, it only needs to be done once.“Single point calibration”
SO-E2-960 of Zirconia Oxygen Sensor for 4. Oxygen Making MachineCharacteristic data:
| Measuring gas | Oxygen |
| Measuring medium | Gas |
| Measuring principle | limiting current type sensor |
| Measurement range | 1%~96% |
| Response time (t90) | 2~25 seconds (depending on sensor type, air flow, measuring chamber) |
| Sensor voltage | 0,7 to 1,6 volts |
| Heating voltage | 3.6 to 4.4 volts |
| Power consumption | 1.3 to 1.8 watts (depending on application and package) |
| Cold resistance | R(25 °c) = 3.25Ω ± 0.20Ω |
| Preheating time | At least 30s. |
| Working temperature | 350°C depending on cable and filter assembly |
| Permissible flow rate [ml/min] | 100~500 (Best 250) |
| Lifetime (MTTF) | 20.000 hours (*) |
Previous Page: O2 Sensor-Thread 3/4NPT Probe O2S-FR-T2-34NPTA-XXX
Next page: SO-E2-250 limiting current oxygen sensor